Abbott Architect Plus ci4100 Chemistry Analyzer Processing Module (Integrated)
This unit is in working condition. Includes everything pictured.
It retails $150k brand new**
Description
The ARCHITECTci4100 combines ARCHITECTc4000 and ARCHITECTi1000SR instruments offering lower volume laboratories the enhanced efficiency and productivity employed by larger volume laboratories.
The ARCHITECTci4100 offers a maximum throughput of up to 900 tests per hour, including 800 clinical chemistry and 100 immunoassay tests. Featuring a load-up capacity of 180 samples with 35 priority positions, the ARCHITECTci4100 has up to 115 refrigerated reagent positions plus Integrated Chip Technology (Na+, K+ and Cl-).
Features
Forin vitrodiagnostic use only.
Specifications
Methods c4000: Photometric, Potentiometric, Turbidimetric; i1000SR:CHEMIFLEX
Maximum Throughput: Up to 900 tests/hour
Sample Types
c4000: Serum, Plasma, Whole Blood, Urine, CSF;i1000SR:Serum, Plasma, Whole Blood, Urine
Sample Tubes: Height: 72102 mm Diameter: 9.616.1 mm
Sample Cup: Yes (50 L dead volume)
Sample Capacity: 180
Sample Barcode Types: Code 39, Codabar, Interleaved 2 of 5, Code 128
Sample Result Storage: 50,000
Sample Volume c4000: 1.535 L Average: 6 L;i1000SR:10150 L Average: 62 L
Automatic Dilution: Yes
Sample Probe Carryover: 0.1 parts per million
Reagent Capacity: Up to 115 refrigerated positions plus patented ISE (Na+, K+, and Cl-)
Reagent Type c4000: >95% liquid ready-to-use; i1000SR: >100% liquid ready-to-use
Reagent Onboard Stability: c4000:565 days Average: 40 days;i1000SR:14-30 days
Calibration Frequency c4000:160 days Average: 25 days;i1000SR: Calibrate with new lot number, if controls are out of range or if specified otherwise within the package insert
Sample, Clot and Bubble Detection: Yes
Reagent Pressure Monitoring: Yes
Sample Interference Measurement: Yes, hemolysis, icterus, and lipemia
System Control Center: 1 SCC, with color touchscreen monitor, keyboard, and mouse
Onboard Maintenance Records: Yes
Online Error Code Help: Yes
Host Interface: Bidirectional, serial RS-232 interface, host query option available
Remote Diagnostics: AbbottLink
Dimension (H x W x D): 49 x 111 x 36 125.1 x 281.2 x 90.7 cm
Weight: 1,677 lbs 760.7 Kg
Electrical Requirements: c4000:AC 180-264V, 47-63 Hz, 20 amp;i1000SR:AC 110-120V or 200-240V, 10%, 50 or 60 Hz self-adjusting
Water Requirements c4000:Deionized water 15 liters/hour during normal operation, 25 liters/hour maximum;i1000SR:Purified water to dilute buffer concentrate
Heat Output*c4000:3050 BTU/hr, running mode;i1000SR:2400 BTU/hr, running mode
Sample Loading: RSH
**Values provided represent the typical output in Running mode for the system processing module and sample handler. A maximum value of 389 BTU/hr was obtained for the System Control Center (SCC) using the ARCHITECTi2000SR as a representative system. Values obtained during an internal study. Data is on file with Abbott Diagnostics.
Additional Prep Fees May Apply